############################### GE 3T Keck Protocol Development ############################### Updates in progress... UTE Spine Protocol ================== Purpose ^^^^^^^ 1) High contrast UTE imaging of the cartilaginous endplate (CEP) and cortical bone 2) Quantitative UTE imaging of the spine using UTE-T2, UTE-MT, UTE-tricomponent and UTE-multicompartment imaging. .. image:: images/Protocols/Ma_UTE_Spine.jpg :width: 500 .. image:: images/Protocols/Ma_UTE_Spine_Protocol.jpg :width: 500 Protocol Steps ^^^^^^^^^^^^^^ 1. Set up the spine coil. Imaging with feet first and using coil elements 4-6 for all the scans. 2. The imaging planes of all the sequences are labeled in the names, except for the localizers. Positioning ^^^^^^^^^^^ 1. Feet first. 2. Align the top of the anterior superior iliac spine with the central region of coil elements 4 and 5. 3. Put the pad under the knee Physiological Gating ^^^^^^^^^^^^^^^^^^^^ Not applicable Post-Processing Support ^^^^^^^^^^^^^^^^^^^^^^^ Dr. Yajun Ma’s group can provide customized MATLAB code for data processing. Acknowledgement ^^^^^^^^^^^^^^^ Thanks to Dr. Yajun Ma’s group in getting this protocol up and running.